HP 16532A - Performance Tests
3-9. VOLTAGE MEASUREMENT ACCURACY
This test verifies the DC voltage measurement accuracy of the instrument. This procedure uses a dual cur- sor measurement that nullifies offset error, resulting in the following measurement:
± [(1.5% of full scale) + (0.016 x V/div)].
Specification:
±[(1.5% of full scale + offset accuracy) + (0.008 x V/div)]
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part numbers rec- ommended.
|
|
| Recommended |
Equipment Required |
| Critical Specification | Model/Part Number |
|
|
| |
DC Power Supply | HP 6114A | ||
Digital Multimeter | Better than 0.1% accuracy | HP 3458A | |
Cable | 50 | Ω BNC (m to m) | HP 10503A |
Adapter | 50 | Ω BNC (f) to Dual Banana Plug | HP |
(cable to power supply) |
|
|
|
Adapter | 50 | Ω BNC tee (m)(f)(f) | HP |
Blocking Capacitor | 50 Ω BNC (m to f) 0.18 μF, 200 V | HP 10240B | |
|
|
|
|
|
|
|
|