3-12. TIME MEASUREMENT ACCURACY

This test uses a precise frequency source to check the accuracy of time measurement functions.

Specification:

±[(0.005% x Δt) + (2 x 106 x delay setting) + 150 ps]

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.

 

 

Recommended

Equipment Required

Critical Specification

Model/Part Number

 

 

 

Signal Generator

200 MHz, timebase accuracy 0.25 ppm

HP 8656B Opt. 001

Cable

50 Ω BNC (m to m) 48-inch

HP 10503A

Adapter

50 Ω Type N (m) to BNC (f)

HP 1250-0780

 

 

 

 

 

 

Procedure:

1.Set the signal generator to 181.81818 MHz (5.5 ns period) and approximately 600 mV rms.

2.Use the N-to-BNC adapter to connect the cable to the signal generator. Connect the free end of the BNC cable to the channel 1 input.

3.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given:

 

Menu

 

Selection

 

Setting

 

 

 

 

 

 

Channel

 

Input

 

1

 

 

 

Probe

 

1:1

 

 

v/Div

 

400 mV

 

 

Offset

 

0.0 V

 

 

Coupling

 

50

Ω / DC

 

 

s/Div

 

2.00 ns

Trigger

 

Mode

 

Edge

 

 

Source

 

Chan 1

 

 

Level

 

0.0 V

 

 

Slope

 

Positive

Display

 

Mode

 

Normal

 

 

Grid

 

On

 

 

 

 

 

 

 

 

 

 

 

 

 

3-13

Image 39
Contents HP 16532A Safety Symbols Safety ConsiderationsContents Service CalibrationSection Replaceable Parts List of Tables Page Table of Contents Page Safety Requirements IntroductionModule Covered by This Manual Operating Characteristics Accessories SuppliedAccessories Available SpecificationsVertical dc Gain Accuracy4 ± 1.5% of full scale HP 16532A SpecificationsRise Time2 1.4 ns Waveform Record Length 8000 pointsHP 16532A Operating Characteristics Vertical at BNCTimebase Trigger Modes TriggeringDigitizer Trigger Level RangeDisplay Modes Waveform DisplayMeasurement Aids Operating Environment Setup AidsEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Preparation for Use Power RequirementsInitial Inspection Safety Requirements Module InstallationProcedure Installation ConsiderationsEndplate Overlap Bottom Sequence Packaging Tagging for ServiceOperating Environment StorageTable of Contents Page Performance Test Procedures Test RecordPerformance Test Interval DC CAL Output Test AC CAL Output Test Specification Equipment Required Critical Specification Model/Part NumberInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Offset Accuracy Procedure Zero-Input OffsetDiv Offset Supply Min Max DC Input OffsetBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Protection Switch Calibration IntervalMaintaining System Integrity Calibration Instrument WarmupLocation of the PROTECT/UNPROTECT Switch Setting the PROTECT/UNPROTECT Switch to UnprotectEquipment Required for Calibration Loading the Default Calibration FactorsCalibrate Delay of the Self Cal menu as follows Self Cal Menu CalibrationsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as followsExt Trig Null Calibration Connections Calibrate the Ext Trig Null of the Self Cal menu as followsSealing the Mainframe Setting the PROTECT/UNPROTECT Switch to ProtectTable of Contents Page Exchange Assemblies Replaceable Parts ListAbbreviations Ordering InformationDirect Mail Order System Reference Designator and Abbreviations Parts Identification Mfr Code Mfr Part Number Reference DesignatorHP Part Qty Number DescriptionPage Table of Contents Page Attenuator Theory of Operation Introduction Safety RequirementsModule Block Diagram and The- ory of Operation Main Assembly Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Tests Self-Test Access ProcedureLoad Test System Screen Touch the Functional Tests field Functional Tests ScreenA/D Test Run Screen Exit Test System Screen Test Self-Test DescriptionsData Memory Test Timebase TestTroubleshooting Aids TroubleshootingVerifying the System Clock Page Troubleshooting E E T Troubleshooting Sheet1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal Procedure Oscilloscope Module Removal and Replacement12. Endplate Overlap Bottom Sequence Oscilloscope Module Replacement ProcedureTrigger Cable Removal Procedure Trigger Cable Removal and ReplacementAttenuator Removal and Replacement Trigger Cable Replacement ProcedureAttenuator Removal Procedure Attenuator Replacement Procedure Warranty CertificationAssistance