3-12. TIME MEASUREMENT ACCURACY
This test uses a precise frequency source to check the accuracy of time measurement functions.
Specification:
±[(0.005% x Δt) + (2 x 10− 6 x delay setting) + 150 ps]
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.
|
| Recommended |
Equipment Required | Critical Specification | Model/Part Number |
|
|
|
Signal Generator | 200 MHz, timebase accuracy 0.25 ppm | HP 8656B Opt. 001 |
Cable | 50 Ω BNC (m to m) | HP 10503A |
Adapter | 50 Ω Type N (m) to BNC (f) | HP |
|
|
|
|
|
|
Procedure:
1.Set the signal generator to 181.81818 MHz (5.5 ns period) and approximately 600 mV rms.
2.Use the
3.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given:
| Menu |
| Selection |
| Setting |
|
|
|
|
|
|
Channel |
| Input |
| 1 |
|
|
| Probe |
| 1:1 | |
|
| v/Div |
| 400 mV | |
|
| Offset |
| 0.0 V | |
|
| Coupling |
| 50 | Ω / DC |
|
| s/Div |
| 2.00 ns | |
Trigger |
| Mode |
| Edge | |
|
| Source |
| Chan 1 | |
|
| Level |
| 0.0 V | |
|
| Slope |
| Positive | |
Display |
| Mode |
| Normal | |
|
| Grid |
| On |
|
|
|
|
|
|
|
|
|
|
|
|
|