HP 1GSa/s Digitizing Oscilloscope 16532A manual Oscilloscope Module Replacement Procedure

Page 85

HP 16532A - Service

5.If a trigger cable was determined to be faulty, replace the faulty trigger cable on the oscilloscope module (refer to paragraph 6-10). Otherwise, skip to step 6.

6.If an attenuator was determined to be faulty, replace the faulty attenuator on the oscilloscope mod- ule (refer to paragraph 6-11). Otherwise, skip to step 7.

7.If the oscilloscope module was determined to be faulty, remove the attenuators from the oscillo- scope module (refer to paragraph 6-11).

8.Acquire a replacement oscilloscope module and install the attenuators on the replacement oscillo- scope module (refer to paragraph 6-11).

9.Reinstall the repaired (or replacement) oscilloscope module in the mainframe (refer to the next pro- cedure "Oscilloscope Module Replacement Procedure").

Oscilloscope Module Replacement Procedure

1.If you have not already done so, pull the installed boards half way out of the card slots (refer to steps 1 through 3 of the preceding procedure "Oscilloscope Module Removal").

2.Determine which modules you want to install in the mainframe, and the slots where you want them installed.

3.Starting with the bottom slot, slide each module to be installed approximately halfway into the main- frame card slot.

4.Firmly seat the bottom card into the backplane connector. Keep applying pressure to the center of the card endplate while tightening the thumb screws finger tight.

5.Repeat step 4 for all cards and filler panels in a bottom-to-top order. See figure 6-12.

Figure 6-12. Endplate Overlap - Bottom Sequence

6.Any filler panels that are not used should be kept for future use. Filler panels must be installed in all unused card slots for correct air circulation.

6-19

Image 85
Contents HP 16532A Safety Symbols Safety ConsiderationsContents Calibration ServiceSection Replaceable Parts List of Tables Page Table of Contents Page Introduction Safety RequirementsModule Covered by This Manual Accessories Available Accessories SuppliedSpecifications Operating CharacteristicsRise Time2 1.4 ns HP 16532A SpecificationsWaveform Record Length 8000 points Vertical dc Gain Accuracy4 ± 1.5% of full scaleVertical at BNC HP 16532A Operating CharacteristicsTimebase Digitizer TriggeringTrigger Level Range Trigger ModesDisplay Modes Waveform DisplayMeasurement Aids Operating Environment Setup AidsEquipment Critical Specifications Recommended Model USE Page Table of Contents Page Initial Inspection Safety Requirements Power RequirementsModule Installation Preparation for UseProcedure Installation ConsiderationsEndplate Overlap Bottom Sequence Operating Environment Tagging for ServiceStorage PackagingTable of Contents Page Test Record Performance Test ProceduresPerformance Test Interval DC CAL Output Test AC CAL Output Test Equipment Required Critical Specification Model/Part Number SpecificationInput Resistance Page Voltage Measurement Accuracy Voltage Measurement Accuracy Test Div Offset Supply Upper Limit Lower Limit Offset Accuracy Procedure Zero-Input OffsetDiv Offset Supply Min Max DC Input OffsetBandwidth V1MHz Time Measurement Accuracy Menu Selection Setting Trigger Sensitivity RUN Repetitive Performance Test Record Paragraph No Test Results Yes No Page Table of Contents Page Calibration Interval Calibration Protection SwitchMaintaining System Integrity Calibration Instrument WarmupLocation of the PROTECT/UNPROTECT Switch Setting the PROTECT/UNPROTECT Switch to UnprotectEquipment Required for Calibration Loading the Default Calibration FactorsSelf Cal Menu Calibrations Calibrate Delay of the Self Cal menu as followsCalibrate Vert Cal of the Self Cal menu as follows Calibrate the Logic Trigger of the Self Cal menu as follows Calibrate the Time Null of the Self Cal menu as followsExt Trig Null Calibration Connections Calibrate the Ext Trig Null of the Self Cal menu as followsSealing the Mainframe Setting the PROTECT/UNPROTECT Switch to ProtectTable of Contents Page Abbreviations Replaceable Parts ListOrdering Information Exchange AssembliesDirect Mail Order System Reference Designator and Abbreviations Parts Identification HP Part Qty Number Reference DesignatorDescription Mfr Code Mfr Part NumberPage Table of Contents Page Module Block Diagram and The- ory of Operation Introduction Safety RequirementsMain Assembly Theory of Operation Attenuator Theory of OperationHP 16532A Oscilloscope Block Diagram Page Self-Tests Self-Test Access ProcedureLoad Test System Screen Touch the Functional Tests field Functional Tests ScreenA/D Test Run Screen Exit Test System Screen Data Memory Test Self-Test DescriptionsTimebase Test TestTroubleshooting Aids TroubleshootingVerifying the System Clock Page Troubleshooting Sheet Troubleshooting E E T1 6 5 3 2 a Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Troubleshooting E E T Oscilloscope Module Removal Procedure Oscilloscope Module Removal and Replacement12. Endplate Overlap Bottom Sequence Oscilloscope Module Replacement ProcedureTrigger Cable Removal Procedure Trigger Cable Removal and ReplacementTrigger Cable Replacement Procedure Attenuator Removal and ReplacementAttenuator Removal Procedure Attenuator Replacement Procedure Certification WarrantyAssistance