HP 16532A - Performance Tests
Procedure:
1.Using a BNC adapter, connect one end of the cable to the power supply. Connect the BNC tee, the blocking capacitor, and the shorting endcap to the other end of the cable. Refer to figure
Figure 3-1. Voltage Measurement Accuracy Test
2.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given.
| Menu |
| Selection | Setting | ||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Channel |
| Input |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1 |
|
| Probe |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1:1 |
|
| Coupling |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1 MΩ / DC |
|
| s/Div |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 500 ns |
Trigger |
| Mode |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Immediate |
Display |
| Mode |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Average |
|
| Average # |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 8 |
|
| Grid |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| On |
|
| Markers |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| On |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Set the Channel Coupling field to 1 MΩ/DC or damage to the equipment will result.