Reference Manual
Rosemount 3051N
When performing a zero trim, ensure that the equalizing valve is open and all wet legs are filled to the correct levels.
NOTE
Do not perform a zero trim on Rosemount 3051N Absolute pressure transmitters. A zero trim is
FIGURE 2-8. Typical Zero vs.
Non-zero-Based Application
Illustrations.
20 mA Point | 20 mA Point |
4 mA Point | 4 mA Point |
LEVEL APPLICATION
Zero Trim
HART Comm. | 1, 2, 3, 3, 1 |
|
|
APPLICATION | APPLICATION |
A full trim is a
To calibrate the sensor with a HART Communicator using the Zero Trim function, perform the following procedure.
1.Vent the transmitter and attach a communicator to the measurement loop.
2.From the communicator menu select 1 Device setup, 2 Diagnostics and service, 3 Calibration, 3 Sensor trim, 1 Zero trim to prepare to adjust the zero trim.
NOTE
The transmitter must be within 3% of true zero
3.Follow the commands provided by the communicator to complete the adjustment of the zero trim.