Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Traditional Flange (H2)

Option PM is a SST pipe mount bracket assembly designed for use in pipe mounting the Traditional Flange (H2). It is the same bracket used on other RNII nuclear qualified transmitters, including the Rosemount 1154 Series H.

When installing the transmitter to the optional mounting brackets, torque the bolts to 21 foot-pounds. See “Mounting” in Section 3 for additional mounting considerations.

The traditional flange option converts the mounting configuration of the Rosemount 3051N to one similar to traditional style transmitters. This allows the Rosemount 3051N to replace traditional transmitters without changing existing manifolds, impulse piping, or bracket arrangements. The traditional flange also allows a higher process temperature at the process ports (300 °F [149 °C]) because of its ability to dissipate heat.

The traditional flange fits most existing mounting brackets. If a new bracket is required, use one of the bracket options described earlier. Figure 6-3 in Section 6 shows the traditional flange.

FIGURE 6-3. Typical Traditional Flange.

TRADITIONAL FLANGE

TRADITIONAL FLANGE WITH DRAIN/VENTS

AND ROSEMOUNT 3051N

 

TRANSIENT PROTECTION TERMINAL BLOCK (T1)

The standard transient protection terminal block increases the Rosemount 3051N Pressure Transmitter’s ability to withstand electrical transients. Rosemount 3051N Pressure Transmitters, with integral transient protection installed, meet the standard performance specifications as outlined in this product manual.

NOTE

Installation of the transient protection terminal block does not provide transient protection unless the Rosemount 3051N transmitter case is properly grounded. See Section 3 for grounding information.

6-8

Image 74
Contents Reference Manual Important Notice -- Errata Rosemount 3051N Return of Material Changes Changes from June 2006 to JuneOld Rosemount 3051N Table of Contents Specifications Reference Data InstallationTroubleshooting Hart Communicator OptionsAppendix a Glossary Index Using this Manual Section IntroductionRosemount 3051N Overview Safety Messages Transmitter FunctionsSecurity Alarm Level VerificationTransmitter Configuring Transmitter Alarm and Security Jumper Procedure Physical Removal Local Zero and Span Local KeysElectronics Board Setting the Loop to Manual Wiring Diagrams Bench Hook-up Wiring Diagrams Field Hook-up Process Variables Review Configuration DataCheck Output Set Process Variable Units Sensor TemperatureBasic Setup Set OutputRerange Rerange with a Communicator OnlyRerange with a Pressure Input Source and a Communicator Span and Zero Adjustment Buttons Detailed Setup Diagnostics ServiceCalibration Transmitter TestLoop Test Transmitter Bench Calibration Tasks Field Calibration Tasks Configure the Analog Output Parameters Calibration OverviewSensor Trim URVZero Trim Typical Zero vs Non-zero-Based Application IllustrationsFull Trim Digital Trim Connection Drawing 4-20 mA TransmittersRecall Factory Trim- Analog Output Digital-to-Analog Trim Using Other ScaleRecall Factory Trim Recall Factory Trim- Sensor Trim Analog Output TrimZero Effect HTc = URV + S URV Ps Rosemount 3051N Section Installation Rosemount 3051N Typical Installation Flowchart Mechanical Considerations General ConsiderationsSpecial Draft Range Considerations 156 181 16 x 11/4 Bolts for Mounting to 54.1 Transmitter 120 127 66.8 Traditional Flange Option Code H2 Dimensional Drawing Mounting Rosemount 3051N GAS or Liquid Service GAS Service Steam Service Electrical Considerations Process Connections Housing RotationPower Supply Wiring Power Supply Load LimitationsEnvironmental Considerations Access Requirements Cover InstallationGrounding the Transmitter Case Troubleshooting Symptom Corrective Actions Returning Rosemount Products and MaterialsNuclear Specifications SeismicModel Range Code Hydrostatic Test Pressure Reference Accuracy Performance SpecificationsAmbient Temperature Effect12 DriftRange Code Overpressure Effect Power Supply Effect Load Effect Mounting Position EffectOverpressure Effect Static Pressure Effect Range Code Static Pressure Zero Effect1Temperature Limits Functional SpecificationsService OutputRange Code Minimum Time Constant Tc Including Dead Time Td Span and Zero, Zero Elevation, SuppressionRange Minimum Span Upper URL Lower LRL Static Pressure Limits Maximum WorkingPressure12 Overpressure Limits Burst PressurePhysical Specifications Materials of Construction Option Code Description Add Ordering Information Mounting Bracket Options Configuration InformationHousing Material Conduit Entry Size Meters optionalDate  On Rosemount 3051N Exploded View with Coplanar Process Flange Rosemount 3051N Rosemount 3051N Overview Safety Messages LCD Meter Section OptionsCustom Meter Configuration Installing the Meter Exploded View of the Rosemount 3051N with Optional LCD MeterRosemount 3051N Diagnostic Messages ErrorPress Limit Operation Mounting BracketsTraditional Flange H2 Transient Protection Terminal Block T1Transient Protection Terminal Block T1 Rosemount 3051N Introduction Safety Messages Appendix a Hart CommunicatorRosemount 3051N Online Menu Function Hart Fast Key Sequence Figure A-3. Rosemount Hart terminal access door Connections and HardwareFigure A-4. Bench Hook-up MA Transmitters Action Keys RosemountCommunicator Keys Data Entry Function KeysAlphanumeric and Shift Keys Menus Functions Fast Key SequencesFast Key Sequence Conventions Fast Key Sequence Example Main MenuOnline Menu Touch Touch Screen Alphanumeric KeysMultifunction LED Multifunction LED Processing IndicationRosemount 3051N Message Description Diagnostic MessagesFast Key Sequences Hot Key options Rosemount 3051N Uploadvariable Rosemount 3051N Rosemount 3051N Glossary Smart Family Index 00809-0100-4808, Rev CA June Security Page ¢00809-0100-4801V¤