Reference Manual
Rosemount 3051N
FIGURE 3-1. Typical
Installation Flowchart.
START HERE
Bench
Calibration? No
Yes
Configure (Section 2)
Set Units
Set Range
Points
Set Output
Type
Set Damping
Verify
Apply Pressure
Within
Specifica- Yes
tions?
No
Refer to
Section 4
Troubleshooting
Field Install
Check Jumpers
and Switches
(page
Mount
Transmitter
(pages
Wire Transmitter (pages
Power
Transmitter
(page
Check Process
Connection
for Leaks
(page
Confirm
Transmitter
Configuration
(page
Trim Transmitter
for Mounting
Effects
(page
Done