Rosemount 3051N
Reference Manual
Overpressure Effect
Static Pressure Effect
Maximum zero shift after overpressure equal to Maximum Working Pressure
Rosemount 3051ND
Range Code | Overpressure Effect |
| |
|
|
|
|
0 | – 3 | ± 0.5% URL |
|
|
|
|
|
4 | – 5 | ± 3.0% URL |
|
|
|
| |
Rosemount 3051NG |
| ||
|
|
| |
Range Code | Overpressure Effect |
| |
|
|
|
|
2 | – 4 | ± 0.25% URL |
|
|
|
|
|
| 5 | ± 0.30% URL |
|
|
|
| |
Rosemount 3051NA |
| ||
|
|
| |
Range Code | Overpressure Effect |
| |
|
|
|
|
0 | – 4 | ± 0.05% URL |
|
|
|
|
|
Rosemount 3051ND
Zero Error (can be calibrated out at line pressure) Per 1000 psi (6,9 MPa) line pressure
Range Code | Static Pressure Zero Effect(1) |
0(2) | ± 0.125% URL for Ps ≤ 750 psi (5,2 MPa) |
1 | ± 0.25% URL for Ps ≤ 2000 psi (13,8 MPa) |
|
|
2, 3 | ± 0.05% URL for Ps ≤ 2000 psi (13,8 MPa) |
| ± [0.1 + 0.1 |
| Ps > 2000 psi ≤ 3626 psi |
|
|
4, 5 | ± 0.1% URL for Ps ≤ 2000 psi (13,8 MPa) |
±[0.2 + 0.2 (Ps-2000) / 1000]% URL for Ps > 2000 psi ≤ 3626 psi
(1)Ps equals static line pressure applied
(2)Specification for Rosemount 3051N Range 0 is expressed in [% per 100 psi (689 KPa)]
Rosemount 3051ND
Span Error
Per 1000 psi (6,9 MPa) line pressure
Range Code | Static Pressure Span Effect |
|
|
0(1) | ± 0.15% input reading |
1 | ± 0.40% input reading |
|
|
2, 3 | ± 0.10% input reading |
|
|
4, 5 | ± 0.20% input reading (uncertainty after calibration correction for |
| systematic effects) |
|
|
(1)Specification for Rosemount 3051N Range 0 is expressed in [% per 100 psi (689 KPa)] up to 750 psi (5 171 KPa)
Power Supply Effect
Load Effect
Mounting Position Effect
Less than ±0.005% of calibrated span per volt for RDF ≤ 10
No load effect other than change in voltage supplied to the transmitter
Rosemount 3051ND/NG
Zero shifts up to ± 1.25 in H20 (0,31 KPa), which can be calibrated out; no span effect
Rosemount 3051NA
Zero shifts up to 2.5 in H20 (63,5 mm), which can be calibrated out; no span effect