Reference Manual
Rosemount 3051N
PERFORMANCE SPECIFICATIONS
Reference Accuracy
Based upon
Includes hysteresis,
Rosemount 3051ND
Range Code | Reference Accuracy |
|
|
0± 0.10% calibrated span from 1:1 to 2:1 RDF
±0.05% upper range limit from 2:1 to 30:1 RDF
1± 0.10% calibrated span from 1:1 to 15:1 RDF
±(0.005% URL + 0.025% span)
| from 15:1 to 50:1 RDF |
|
|
2 – 5 | ± 0.075% calibrated span from 1:1 to 10:1 RDF |
| ± (0.005% URL + 0.025% span) from 10:1 to 100:1 RDF |
|
|
Rosemount 3051NG | |
|
|
Range Code | Reference Accuracy |
|
|
2 – 5 | ± 0.075% calibrated span from 1:1 to 10:1 RDF |
| ± (0.005% URL + 0.025% span) from 10:1 to 100:1 RDF |
|
|
Rosemount 3051NA
| Range Code | Reference Accuracy |
| |
|
|
|
|
|
|
| 0 | ± 0.075% calibrated span from 1:1 to 5:1 RDF |
|
|
|
| ± (0.01% URL + 0.025% span) from 5:1 to 30:1 RDF |
|
|
|
|
|
|
| 1 | – 4 | ± 0.075% calibrated span from 1:1 to 10:1 RDF |
|
|
|
| ± (0.0075% URL) from 10:1 to 100:1 RDF |
|
|
|
|
| |
Drift | Rosemount 3051ND, NG, NA |
| ||
|
|
|
| |
| Range Code | Drift |
| |
|
|
|
|
|
|
| 1 | ± (0.2% URL + 0.2% span) for 30 months |
|
| 2 | – 5 | ± 0.2% URL for 30 months |
|
|
|
|
|
|
Ambient Temperature Effect(1)(2)
Rosemount 3051ND/NG
Range Code | Ambient Temperature Effect per 50° F (28° C) |
|
|
0± (0.25% URL + 0.05% span)
1± (0.1% URL + 0.25% span) from 1:1 to 30:1 RDF
± (0.14% URL + 0.15% span) from 30:1 to 50:1 RDF
2 – 5 | ± (0.0125% URL + 0.0625% span) from 1:1 to 5:1 |
| ± (0.025% URL + 0.125% span) from 5:1 to 100:1 |
Rosemount 3051NA
Range Code | Ambient Temperature Effect per 50° F (28° C) | ||
|
|
| |
0 |
| ± (0.1% URL + 0.25% span) | |
|
|
|
|
1 – | 4 | ± (0.025% | URL + 0.125% span) from 1:1 to 30:1 |
|
| ± (0.035% | URL + 0.125% span) from 30:1 to 100:1 |
|
|
|
|
(1)RDF = Range Down Factor = URL / Calibrated Span
(2)Exposure of isolating diaphragms to process temperatures above 185 °F (85 °C), but below 250 °F (121 °C), produces a temperature effect of ±1.0% of calibrated span in addition to the effects listed.