Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

Glossary

Analog Output Trim

Cloning

Commissioning

Configuration

Damping

Descriptor

Digital Trim

Failure Mode Alarm

Factory Characterization

Full Trim

HART (Highway

Addressable Remote

Transducer) Protocol

Lower Range Limit (LRL)

Digital Trim operation that allows adjustment of the output electronics to conform to the plant standard. Three types of analog output trim are available: 4–20 mA Trim, 4–20 mA Other Scale, and Low Power. See “Analog Output Trim” in Section 2 . Low Power is not approved for use with the Rosemount 3051N.

Off-line operation that uses the HART Communicator (Rosemount 275 or Rosemount 375) to copy configuration data from one transmitter to one or more other transmitters that require the same data.

Functions performed with a HART-based communicator and the transmitter which test the transmitter, test the loop, and verify transmitter configuration data. See “Commissioning the Rosemount 3051N with a HART-Based Communicator” in Section 2 .

Process of setting parameters that determine how the transmitter operates.

Output function that increases the response time of the transmitter to smooth the output when there are rapid input variations. See “Damping” in Section 2 .

Sixteen-character field for additional identification of the transmitter, its use, or location. The descriptor is stored in the transmitter and can be changed using the HART-based communicator.

Format function that allows you to adjust the transmitter characterization for purposes of digital calibration to plant standards. Digital Trim includes two separate operations: Sensor Trim and Analog Output Trim. See “Transmitter Functions” in Section 2.

Transmitter function that drives the analog output to a jumper-selectable high or low value in the event of an electronics failure. See “Failure Mode Alarm” in Section 2 .

Factory process during which each sensor module is subjected to pressures and temperatures covering the full operating range. The sensor module memory stores data generated from this process for use by the microprocessor in correcting the transmitter output during operation.

Sensor Trim function in which two accurate, end-point pressures are applied, and all output is linearized between them. The selected end points should always be equal to or outside the LRV and URV. See “Full Trim” in Section 2 .

Communications standard that provides simultaneous analog and digital signal transmission between control rooms and field devices such as transmitters. All Rosemount SMART FAMILY products communicate using the HART protocol.

Lowest value of the measured variable that the transmitter can be configured to measure.

www.rosemountnuclear.com

Image 95
Contents Reference Manual Important Notice -- Errata Rosemount 3051N Return of Material Changes Changes from June 2006 to JuneOld Rosemount 3051N Table of Contents Specifications Reference Data InstallationTroubleshooting Hart Communicator OptionsAppendix a Glossary Index Section Introduction Using this ManualRosemount 3051N Transmitter Functions Overview Safety MessagesSecurity Alarm Level VerificationTransmitter Physical Removal Local Zero and Span Local Keys Configuring Transmitter Alarm and Security Jumper ProcedureElectronics Board Setting the Loop to Manual Wiring Diagrams Bench Hook-up Wiring Diagrams Field Hook-up Process Variables Review Configuration DataCheck Output Set Output Sensor TemperatureBasic Setup Set Process Variable UnitsRerange with a Communicator Only RerangeRerange with a Pressure Input Source and a Communicator Span and Zero Adjustment Buttons Diagnostics Service Detailed SetupCalibration Transmitter TestLoop Test Transmitter Bench Calibration Tasks Field Calibration Tasks Calibration Overview Configure the Analog Output ParametersURV Sensor TrimTypical Zero vs Non-zero-Based Application Illustrations Zero TrimDigital Trim Connection Drawing 4-20 mA Transmitters Full TrimAnalog Output Trim Digital-to-Analog Trim Using Other ScaleRecall Factory Trim Recall Factory Trim- Sensor Trim Recall Factory Trim- Analog OutputZero Effect HTc = URV + S URV Ps Rosemount 3051N Section Installation Rosemount 3051N Typical Installation Flowchart Mechanical Considerations General ConsiderationsSpecial Draft Range Considerations 156 181 16 x 11/4 Bolts for Mounting to 54.1 Transmitter 120 127 66.8 Traditional Flange Option Code H2 Dimensional Drawing Mounting Rosemount 3051N GAS or Liquid Service GAS Service Steam Service Electrical Considerations Process Connections Housing RotationPower Supply Power Supply Load Limitations WiringEnvironmental Considerations Access Requirements Cover InstallationGrounding the Transmitter Case Troubleshooting Returning Rosemount Products and Materials Symptom Corrective ActionsSeismic Nuclear SpecificationsModel Range Code Hydrostatic Test Pressure Drift Performance SpecificationsAmbient Temperature Effect12 Reference AccuracyRange Code Static Pressure Zero Effect1 Power Supply Effect Load Effect Mounting Position EffectOverpressure Effect Static Pressure Effect Range Code Overpressure EffectOutput Functional SpecificationsService Temperature LimitsRange Code Minimum Time Constant Tc Including Dead Time Td Span and Zero, Zero Elevation, SuppressionRange Minimum Span Upper URL Lower LRL Overpressure Limits Burst Pressure Maximum WorkingPressure12 Static Pressure LimitsPhysical Specifications Materials of Construction Option Code Description Add Ordering Information Meters optional Configuration InformationHousing Material Conduit Entry Size Mounting Bracket OptionsDate  On Rosemount 3051N Exploded View with Coplanar Process Flange Rosemount 3051N Rosemount 3051N Section Options Overview Safety Messages LCD MeterCustom Meter Configuration Exploded View of the Rosemount 3051N with Optional LCD Meter Installing the MeterRosemount 3051N Error Diagnostic MessagesPress Limit Mounting Brackets OperationTransient Protection Terminal Block T1 Traditional Flange H2Transient Protection Terminal Block T1 Rosemount 3051N Appendix a Hart Communicator Introduction Safety MessagesRosemount 3051N Online Menu Function Hart Fast Key Sequence Connections and Hardware Figure A-3. Rosemount Hart terminal access doorFigure A-4. Bench Hook-up MA Transmitters Action Keys RosemountCommunicator Keys Data Entry Function KeysAlphanumeric and Shift Keys Main Menu Fast Key SequencesFast Key Sequence Conventions Fast Key Sequence Example Menus FunctionsOnline Menu Touch Multifunction LED Processing Indication Alphanumeric KeysMultifunction LED Touch ScreenRosemount 3051N Message Description Diagnostic MessagesFast Key Sequences Hot Key options Rosemount 3051N Uploadvariable Rosemount 3051N Rosemount 3051N Glossary Smart Family Index 00809-0100-4808, Rev CA June Security Page ¢00809-0100-4801V¤