Omega Engineering 25, 48, 20, 30 Additional information, Fault Finding, Fuses, Insulation testing

Page 35

12. Additional information

Brief fault finding notes and a list of replacement parts are given in this section.

Note that this equipment should only be dismantled by properly trained personnel. Removing the outer cover exposes potentially lethal power voltages.

There are no user serviceable parts within this equipment.

Fault Finding

Should you encounter problems with your TCY contact Omega Flow Engineering for assistance.

Fuses

If neither the power light nor display on the front panel is lit, one of the two fuses may have blown. Check that there is no external cause, such as a faulty plug or lead. Check both fuses and replace the faulty fuse with a new one of the correct value (fuse values are given on the label next to the power inlet). Note that fuses should only be replaced by a qualified electrician.

The holder for the two fuses is built into the power input socket. First remove the power cable and then gently prise the fuse drawer open with a flat-bladed screwdriver or similar tool.

Each fuse can be removed by using the screwdriver as a lever.

Exchange the faulty fuse in the fuse holder for a working fuse of the correct value. Finally, replace the fuse drawer in the fuse compartment and push the drawer shut.

Fuses which blow repeatedly indicate a serious fault and you should return the unit to your supplier for repair.

The heated lid over-temperature cut-out

The heated lid is fitted with an independent circuit to protect it from overheating. In the unlikely event of an over-temperature problem with the lid, the unit is fitted with a thermal fuse which will remove power to the heater plate should the maximum temperature be exceeded.

Insulation testing

This equipment is fitted with RFI suppression circuitry. Any check of the electrical insulation by means of high voltage dielectric testing (for example as in BS EN 61010-1) must be carried out using only a TCY voltage.

This unit contains semiconductor components which may be damaged by electric field effects.

Interchangeable blocks

The block can be removed and replaced by another of the same sort or of a different sort. The software will register the change and set up the unit for the new block.

Remove the four screws on the underside of the unit and drop the block/fan assembly down. Unplug the assembly from the wiring and remove the assembly from the unit. Refitting the block/fan assembly is the reverse procedure; be careful of the wires as you refit the block.

35

Image 35
Contents Shop online at TCY 20, 25, 30Thermal Cyclers User’sGuideFor immediate technical or application assistance OMEGAnet Online Service omega.comInternet e-mail info@omega.com Servicing North AmericaContents Operator Safety 2. Safety and installation1. Introduction ELECTRICIAN InstallationReplacement cable UnpackingWorking Conditions Heated Lid 3. Cycler TCY SpecificationTemperature Heating/Cooling RateHeated lid 4. User operation and consumablesDimensions Power0.5ml Tubes Information Configuration TCYTCProgramsMethods00 to selectMethods Configuration5. Front panel controls ProgramsThe Down arrow key The keysThe End, Finish or Exit key The Up arrow keyThe Alpha/Numeric keys 6. Programming the CYCLERStep MethodsPrograms00 Information ConfigurationStage CycleOption 1 Creating programs using templates 2 STEP TEMPLATESTEP Run ProgramMethodTo save the program To give a program a nameTo edit a field or step in an existing program ProgramsMethods00 Information ConfigurationSet password to select to select DONNA 7 Run programmethod5m00 Post hold10.0 72.0ºC105ºC UNNAMED #10.0ºCºC To set a password for a selected programTypical 3-Step PCR Example programsProgramming New*New- program ** 2 STEP TEMPLATE 5012 3 STEP TEMPLATEdescribed above To insert additional stepsTo delete a step to edit the step asTo add a new stage at the end to insertstage To add a new stage before an existing oneThe prompt line will flash between ‘CHANGE TO PAUSE?’ and ‘YES or NO’ To insert a ‘pause’ into a programto go into edit mode To program increment or decrement temperature or timeTemperature profile of incremented Hold Time at step 3 of each cycle Press “Enter” Press “Pause”twice to go into edit Stopping what you are doing Run ProgrammethodGEORGE View methodProgramTo run a program Run methodProgramNewprogramMethods to select72.0ºC Sample58.0ºC Hold @Unit s/n 8. Configuration7. Information Block s/nidSettings ProgramSettingsdefaultsxxxx PasswordsFormat memoryPasswords Thiswill delete SURE ?9. Notes on transferring protocols from other thermal cyclers Underside of the CYCLER TCY 10. User maintenanceCleaning your CYCLER Heater Block removal11. Some frequently asked questions Fuses 12. Additional informationFault Finding Insulation testingRETURN REQUESTS/INQUIRIES WARRANTY/DISCLAIMERpH/CONDUCTIVITY TEMPERATUREPRESSURE, STRAIN AND FORCE FLOW/LEVEL