3. Cycler TCY Specification
Temperature
Temperature range | 4°C to 99°C |
Temperature set point precision | 0.1°C |
Block uniformity (over full range) | ± 0.5°C |
Temperature accuracy | ± 0.5°C |
End of program cooling below ambient | Yes |
Heating/Cooling Rate
Maximum heating rate | 4.5ºC/s |
Minimum heating rate | 0.1ºC/s |
Maximum cooling rate | 3.0ºC/s |
Minimum cooling rate | 0.1ºC/s |
Heated Lid
Heated lid enable/disable | Yes |
Selectable heated lid temperature | 100°C to 115°C |
Fixed at 145°C | |
Temperature sensor | Thermistor |
Heater type | Etched foil |
Heater power | 33W |
Warm up time | 4 min approximately |
| (ambient to 105°C) |
The heated lid will only come on if the set temperature is above 35°C |
Programming
Program naming using alpha keys | Yes |
Program password protection | Yes |
Number of programs (3 step) | 37 |
Maximum number of steps per stage | 35 |
Maximum number of stages per program | 25 |
Minimum hold time | 1 sec |
Maximum hold time | 18 hrs |
Incremental/decremental temperature | Yes |
Incremental/decremental step hold time | Yes |
Programmable ramp rate (heat or cool) | Variable in 0.1°C/sec steps |
Programmable ramp rate resolution | 0.1°C |
Running Programs
Pause button | Yes |
Stop button | Yes |
End of program alarm | Yes |
Auto restart on power fail | Yes |
Interconnectivity
Connection to PC control program | Yes |
Software updates from | Yes |
7