Working Conditions
The thermal cycler is designed operate under the following conditions:
∙Indoor use
∙Ambient temperature range +5oC to +40oC
∙Altitude to 2000m
∙Relative humidity not exceeding 95%
∙Power supply fluctuations not exceeding 10%
∙Over voltage category II IEC
∙Pollution degree 2
Note: The control specifications are quoted at an ambient temperature of 20oC. The specification may deteriorate outside an ambient temperature of 10oC to 30oC.
The instrument has been tested for radio frequency interference and is certified under EN61326.
6