8
2-2 Installing the Balance
Caution
The weighing unit and the display unit were adjusted as a unit. Therefore, make sure that
the weighing unit and the display unit have the same serial number. The serial number is
printed on the rear of the weighing unit and the display unit. The connection cable also
carries the serial number.
If the serial numbers of both units are different, the balance may not function properly.
When a repair is necessary, submit both the weighing unit and the display unit for repair.
There is no compatibility between The AD-4212A and the AD-4212B or AD-4212, for both the
weighing unit and the display unit.
Install the balance as follows:
1. Refer to “3. PRECAUTIONS” for installing the balance.
2. Refer to "2-1 Unpacking" on the previous page to attach the dust guard, the breeze ring
(AD-4212B-101/102 only), the pan support and the weighing pan on the weighing unit.
The stainless steel breeze break is provided for the AD-4121B series balance and the
AD-4212A-100 as a standard accessory, and for the other models as an option. Use it as
necessary when performing calibration or checking accuracy.
How to install the breeze break
AD-4212A-100/200 and the AD-4212B series balance
Place the breeze break on the weighing pan so that it fits over the dust guard.
AD-4212A-600/1000
Follow the procedure below to install the breeze break.
1 Remove the weighing pan and the pan support.
2 Place the breeze break on the weighing pan so
that it fits over the dust guard.
3 Replace the pan support.
4 Replace the weighing pan.
3. Connect the weighing unit and the display unit, firmly inserting one end of the connection cabl e
into the jack located on the rear of the weighing unit and the other end into the jack located on
the rear of the display unit.
If the extension cable (OP-07: 3 m) is used, connect it between the connection cable and the
display unit.
Pan support
2
2
3
4
Weighing
pan
Breeze
break
Dust guard