36

Key mode B

When the PRINT key is pressed or the external PRINT contact input is used, the balance outputs or
stores the weighing data immediately regardless of the display condition. At this time, the display
does not blink.

Key mode C

When the PRINT key is pressed or the external PRINT contact input is used, with the stabilization
indictor on, the balance outputs or stores the weighing data immediately. If the stabilization indicator
is not on, the balance waits for the indicator to turn on, and then outputs the weighing data. At this
time, the display blinks one time.

Interval memory mode

The weighing data is periodically stored in memory.
Required setting dout prt 3 Interval memory mode
dout data 2 Data memory function is used.
Stores weighing data.
dout int Interval time
Optional setting dout 5-td1, 2, or 3 Adds the time and date.
Example “For periodical weighing without a computer command and to output
all of the data, to a computer, at one time”
10-6 Description of the Item “Data format”

A&D standard format 5if type 0

This format is used when the peripheral equipment can receive the A&D format.
If an AD-8121B is used, set the printer to MODE 1 or 2.
This format consists of fifteen or sixteen characters excluding the terminator.
When numerical characters excluding a decimal point are exceeded eight characters for
AD-4212B-102, the format becomes sixteen characters.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zer os. If the data is zero, the plus
sign is applied.
The unit, consisting of three characters, follows the data.
When comparison results are to be added (Cp fnc Cp-r 1), the results appear between the
header and the data.
CRLF
ST,+000.1278 g
ST
US
OL
QT
CRLF
ST,+100.01278 g
Header Data (eight characters) Unit Terminator
Stable header Stable header of counting mode
Unstable header
Overload header
Header Data Unit Terminator