35

Tare at start ( p-tr )

When a hopper is attached to the weighing pan and loss-in weighing is performed, the remaining
amount of the material will become unknown if tare is performed each time a weighing starts.
When "p-tr 0" is selected, tare is not performed at weighing start. So, the remaining amount of the
material can be monitored when the power is turned on.
10-5 Description of the Item “Data output mode”
The parameter setting of the “Data output mode ( prt )” applies to the performance when the “Data
memory ( data )” parameter is set to “2” (to store the weighing data) and when the data is transmitted
using the RS-232C interface.
Notes
Data output to the RS-232C is not available when the BCD output (OP-01) is installed.
BCD output is refreshed at the display refresh rate.

Key mode

When the PRINT key is pressed or the external PRINT contact input is used, with the stabilization
indictor on, the balance outputs or stores the weighing data and the display blinks one time.
Required setting dout prt 0 Key mode

Auto print modes A and B

When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”
and reference value are met, the balance outputs or stores the weighing data.
When the PRINT key is pressed with the stabilization indictor on, the balance outputs or stores the
data and the display blinks one time.
Mode A: Required setting dout prt 1 Auto print mode A (reference = zero)
dout ap-p Auto print polarity
dout ap-b Auto print difference
Example “With ar-d 1” (to adjust zero after data is output), for weighing the
added sample each time a sample is added.”
Mode B: Required setting dout prt 2 Auto print mode B (reference =
last stable value)
dout ap-p Auto print polarity
dout ap-b Auto print difference
Example “For weighing while a sample is added.”

Stream mode

The balance outputs the weighing data continuously regardless of the display condition. The display
does not blink in this mode. This mode is not available and the interval memory mode is used when
the “Data memory ( data )” parameter is set to “2” (to store the weighing data).
Required setting dout prt 3 Stream mode
dout data 0 Data memory function is not used.
ba5fnc 5pd Display refresh rate
5if bp5 Baud rate
Example “For monitoring data on a computer”

Caution

The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate or data added to the weighing data such as time, date or ID number.