48
Calibration test report using an external weight
(Calibration test does not perform calibration.)
When the setting is “info 1”: When the setting is “info 2”:

AD-8121 format General data format

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration test

type

Zero point value

Targ et wei ght

value

Targ et wei ght

Signature

CAL.TEST(EXT.)
ACTUAL
0.00000 g
+50.00020 g
TARGET
+50.00000 g
SIGNATURE
- - - - - - - -
A&D
 
S/N
ID
DATE
TIME
AD4212B-102
01234567
ABCDEFG
2000/12/31
12:34:56
A & D <TERM>
AD4212B-102<TERM>
S/N 01234567<TERM>
ID ABCDEFG<TERM>
DATE<TERM>
200 0/12/ 31<TERM>
TIME<TERM>
12:34:56<TERM>
CAL.TEST(EXT.)<TERM>
ACTUAL<TERM>
0.00000 g<TERM>
+50.00020 g<TERM>
TARGET<TERM>
+50.00000 g<TERM>
SIGNATURE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
<TERM> CRLFCR

Space, ASCII 20h

Terminator, , or

Carriage return, ASCII 0Dh

Line feed, ASCII 0Ah

CR
LF