30
Class Item and Parameter Description
0
3-level comparator contact
output
Cpout
Comparator
output
Cp-t
Comparator output
setting
1
5-level comparator contact
output
Only the AD-4212B series
balance displays this class and
item.
0
No comparison
1 Comparison, excluding “near zero” when stable or overloaded
2 Comparison, including “near zero” when stable or overloaded
3 Continuous comparison, excluding “near zero”
Cp
Comparator mode
4 Continuous comparison, including “near zero”
0 Not added
Cp-r *2
Comparison results
1 Added Use in A&D standard format.
0 OFF bep
LL buzzer 1
ON
Displayed only when 5-level
comparator output (*3) is selected.
0
OFF
bep_
LO buzze
r
1
ON
0
OFF
bep-
OK buzze
r
1
ON
0 OFF bep
HI buzzer 1
ON
0 OFF
Cp fnc
Comparator
bep
HH buzzer 1 ON
Displayed only when 5-level
comparator output (*3) is selected.
Cp HH
Secondary upper limit
Cp Hi
Upper limit
Cp lo
Lower limit
Cp ll
Secondary lower limit
Refer to "10-10 Comparator
Function”.
Displays Cp HH,, Cp ll only when
5-level comparator output (*3) is
selected.
0 Key mode (when stable)
Accepts the PRINT key or external
contact input to output (or store)
data only when the display is
stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs (or stores) data when the
display is stable and conditions of
ap-p, ap-b and the reference value
are met.
3
Stream mode /
Interval memory mode
With data 0, outputs data
continuously; with data 2, uses
interval memory.
4 Key mode B (instantly)
Accepts the PRINT key or external
contact input to output (or store)
data, regardless of the display
condition.
dout
Data output
Not displayed
when BCD
output (OP-01)
is installed.
prt
Data output mode
5 Key mode C (when stable)
Accepts the PRINT key or external
contact input to output (or store)
data, instantly when stable or after
the display is stabilized.
Factory setting
*2 This item is not displayed for the AD-4212B series balance or when BCD output (OP-01) is
installed.
*3 AD-4212A series balance: when OP-04 is installed
AD-4212B series balance: when “Cp-t” is set to “1