18
6. WEIGHING

6-1 Basic Operation (Gram Mode)

1 Plug in the AC adapter.
2 Press the ON:OFF key to display the weighing mode. (The
decimal point position depends on the balance model.)
Notes
The auto display-ON function is available to display the
weighing mode without the key operation when the AC
adapter is plugged in. (Function table: ba5fnc p-on)
The function not to tare at start can be selected. (Function
table: ba5fnc p-tr)
For details on the function table settings, refer to "10.
FUNCTION TABLE".
The factory setting of the minimum weighing value for the
AD-4212B-101/102 is 0.1 mg. To display 0.01 mg, press the
SAMPLE key. Then, warm up the balance for the
appropriate duration.
3 Place a container on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The
balance displays 0.0000 g. (The decimal point position
depends on the balance model.)
4 Place a sample on the pan or in the container. Wait for the
stabilization indicator to be displayed and read the value.
5 Remove the sample and container from the pan.
Notes
To use another unit, refer to “5-2 Changing the Unit”.
Press the SAMPLE key to switch the minimum weighing
value.
The function to turn on or off the minimum weighing value
automatically when weighing is started can be selected.
(Function table: ba5fnc rng ) For details, refer to "10.
FUNCTION TABLE".
The weighing data can be stored in memory. For details,
refer to “14. DATA MEMORY”.
Weighing
pan
Sample

Remove the

sample

(Software
version
display)
(Display OFF)
(All segments
display)
(Weighing
mode)
Container