45
Note
The upper/lower limit values can be confirmed by external commands using the RS-232C serial
interface.
e.g., Command ?HH (to confirm secondary upper limit value)
Response HH, +100.00000 g (secondary upper limit value=100.00000 g)
Command ?HI (to confirm upper limit value)
Response HI +080.00000 g (upper limit value=80.00000 g)
Command ?LO (to confirm lower limit value)
Response LO, + 060.00000 g (lower limit value=60.00000 g)
Command ?LL (to confirm secondary lower limit value)
Response LL, +040.00000 g (secondary lower limit value=40.00000 g)
indicates a space (20h)
Adding the comparison results (AD-4212A only)
To add the comparison results to the output data, set “Comparison results (Cp-r)” parameter of the
function table to “1”.
ST, HH, 110 .0 000 g HH (Only when OP-04 is installed.)
ST, HI, +090.0000 g HI
ST, OK,+070.0000 g OK
ST, LO, +050.0000 g LO
ST, LL,+030.0000 g LL (Only when OP-04 is installed.)
ST, - -,+000.0000 g No comparison when unstable or near zero
indicates a space (20h)
Notes
Select A&D standard format. Set “Data format (type)” parameter of the function table to “0”.
The comparison results can not be added to the weighing data stored in memory.