9-2 Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
Upper/lower limit values
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
Calibration weight value
Function switch settings
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing an d holding the PRINT key and the
SAMPLE key, press the ON:OFF key. The balance
displays p5 .
3 Press the SAMPLE key to display Clr .
4 Press the PRINT key.
To cancel this operation, press the CAL key.
5 Press the RE-ZERO key.
6 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing
mode.