82
23. SPECIFICATIONS
AD-4212A-100 AD-4212A-200 AD-4212A-600 AD-4212A-1000
Weighing capacity 110 g 210 g 610 g 1100 g
Maximum display 110.0084 g 210.084 g 610.084 g 1100.084 g
Minimum weighing value (1 digit) 0.1 mg 1 mg
Repeatability (Standard deviation) 0 .15 mg 1 mg
Linearity
±
0.3 mg
±
2 mg
±
3 mg
0-30 g 1.1 0-30 g 0.8 0-30 g 0.9 0-30 g 0.9
Stabilization time in seconds
(typical at FAST under good
environment) 3 0-110 g 1.3 30-210 g 1.0 30-61 0 g 1.1 30-1100 g 1.1
Display refresh rate 10 times/second (5 times/second can be selected)
RS-232C Bi-directional, 600-19200 bps
Comparator contact
output 3-level output
I/O unit
External contact input RE-ZERO, PRINT
Buzzer Sounds the connected buzzer for 3-level comparison results
(It becomes 5-level when OP-04 is installed)
Sensitivity drift ±2 ppm/°C (10°C-30°C/50
°
F-86
°
F )
Operating environment 5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation)
Calibration weight provided
(Conforming to OIML Class E2) 50 g 100 g 200 g
Applicable weight values
20 g, 50 g, 100 g 20 g, 50 g,
100 g, 200 g
20 g, 50 g, 100 g,
200 g, 300 g, 400 g,
500 g, 600 g
20 g, 50 g, 100 g, 200 g,
300 g, 400 g, 500 g, 600 g,
700 g, 800 g, 900 g, 1000 g
Dimensions 80 (W)
×
230 (D)
×
90 (H) mm
Weighing pan 50
×
50 mm 70
×
70 mm
Weighing
unit
Mass Approx. 2 kg
Dimensions 237 (W)
×
150 (D)
×
155 (H) mm Display
Mass Approx. 2 kg
Connection cable Approx. 2 m (between weighing unit and display)
AC adapter
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption Approx. 11VA (supplied to the AC adapter )
Minimum unit mass 0.1 mg 1 mg Counting
mode Number of samples 10, 25, 50 or 100 pieces
Minimum 100%
reference mass
10.0 mg 100 mg
Percent
mode
Minimum 100%
display
0.01%
0.1%
1%
(Depends on the reference mass stored.)
OP-01 BCD output, external contact input
RS-232C and comparator contact output disabled
OP-04 RS-232C, 5-level comparator contact output, external contact input
(Not field installable. Must be ordered with the balance for installation at the factory.)
OP-07
*
Extension cable 3 m
OP-08 Ethernet interface
RS-232C and comparator contact output disabled
OP-19 Stainless steel breeze break
Provided as standard for AD-4212A-100
Options
OP-20 Metal leveling foot
Only one extension cable can be used. The maximum distance between the weighing unit
and the display unit is 5 m (connection cable 2 m + extension cable 3 m).