The following commands can not be used during data storage.
Q Query command for weighing data.
S Query command for stable weighing data.
SI Query command for weighing data.
SIR Query command for continuous weighing data.
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode
Auto print
polarity,
difference
Data memory
function
Interval
time
Key mode prt 0, 4, 5Not used data 2
Auto print mode A prt 1 data 2
Auto print mode B prt 2
ap-a 0-2
ap-b 0-2 data 2
Not used
Interval memory mode prt 3 Not used data 2 int 0-8
Set each item, depending on the situation, as follows:
No d-no 0 No 5-td 0
Data number
Yes d-no 1 Time only 5-td 1
No 5-id 0 Date only 5-td 2 ID number
Yes 5-id 1
Time and date
Both 5-td 3
With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.
Confirm that the “Data memory ( data )” parameter is set to “2”.
1 Press and hold the PRINT key until reCall is displayed, then release the key.
2 Press the PRINT key to enter the memory recall mode. Recall the data in memory using the
following keys
RE-ZERO key To proceed to the next data set.
MODE key To go back to the previous data set.
PRINT key To transmit the current data using the RS-232C interface.
With SAMPLE held down, press the CAL key
To delete the data currently displayed.
Note: Deleting the data will not increase the amount of data that can be stored.
CAL key To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.