32
Class Item and Parameter Description
0 600 bps
1 1200 bps
2 2400 bps
3 4800 bps
4 9600 bps
bp5
Baud rate
5 19200 bps
0 7 bits, even
1 7 bits, odd
btpr
Data bit, parity bit
2 8 bits, none
0 CR LF
Crlf
Terminator
1 CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0 A&D standard format
1 DP format
2 KF format
3 MT format
4 NU format
type
Data format
5 CSV format
Refer to "10-6 Description of
Item "Data Format".
0 No limit
t-Up
Timeout
1 1 second
Selects the wait time to receive a
command.
0 No output
erCd
AK, Error code
1 Output AK: ASCII code 06h
0 Not used
5if
Serial
interface
Not displayed
when BCD
output (OP-01)
is installed.
Ct5
CTS, RTS control
1 Used Controls CTS and RTS.
0 OFF at 1 ON at 0
datp
Data
1 ON at 1 OFF at 0
0 OFF at negative ON at positive or zero
polp
Polarity
1 ON at negative OFF at positive or zero
0 OFF at stabilization indicator ON ON at stabilization indicator OFF 5tbp
Stability
1 ON at stabilization indicator ON OFF at stabilization indicator OFF
0 OFF at e, -e display 0erp
OVER
1 ON at e, -e display
0 Data refresh completed when ONOFF
5trp
STROBE
1 Data refresh completed when OFFON
0 OFF when weighing ON when not weighing *4
bcd
Displayed only
when BCD
output (OP-01)
is installed.
5tap
Status
1 ON when weighing OFF when not weighing *4
Unit
Unit
Refer to “5. WEIGHING UNITS”.
id
ID number setting
Refer to “11. ID NUMBER AND GLP REPORT”.
erfnc
Extended functions
Available only for the AD-4212A series balance when “To use
the extended functions” is selected for the function switch and is
used for special applications.
For details, refer to “20. EXTENDED FUNCTIONS”.
Factory setting
*4 “Not weighing” is a condition that a weight value is not displayed such as re-zero operation or
display-off state.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate or data added to the weighing data such as time, date or ID number.