Programming the DC Source - 3
Internally Triggered Measurements
You can use the data acquisition trigger system to synchronize the timing of the voltage and current data acquisition with a BUS or internal trigger source. Then use the FETCh commands to return different calculations from the data acquired by the measurement trigger.
SCPI Triggering Nomenclature
As previously explained under "Triggering Output Changes", the dc source uses the following sequence name and alias for the measurement trigger system. This alias can be used instead of the sequence form.
Sequence Form | Alias |
SEQuence2 | ACQuire |
Measurement Trigger System Model
Figure
ABORt
IDLE STATE *RST *RCL
INITiate[:IMMediate]
INITIATED STATE
TRIGGER RECEIVED
SENSe:SWEep:POINts
ACQUIRED
NO TRIGger:COUNt
COMPLETE?
YES
Figure
Initiating the Measurement Trigger System (Agilent 66312A, 66332A Only)
When the dc source is turned on, the trigger system is in the idle state. In this state, the trigger system ignores all triggers. Sending the following commands at any time returns the trigger system to the Idle state:
ABORt
*RST
*RCL
The INITiate commands move the trigger system from the Idle state to the Initiated state. This enables the dc source to receive triggers. To initiate for a measurement trigger, use:
25