Language Dictionary - 4
TRIGger:SEQuence2:COUNt:CURRent
TRIGger:ACQuire:COUNt:CURRent
Agilent 66312A, 66332A Only
This command sets up a successive number of triggers for measuring current data. With this command, the trigger system needs to be initialized only once at the start of the acquisition period. After each completed measurement, the instrument waits for the next valid trigger condition to start another measurement. This continues until the count has completed.
Command Syntax | TRIGger:SEQuence2:COUNt:CURRent<NRf+> | |
| TRIGger:ACQuire:COUNt:CURRent<NRf+> | |
Parameters | 1 to 100 |
|
*RST Value | 1 |
|
Examples | TRIG:SEQ2:COUN:CURR 5 TRIG:ACQ:COUN:CURR 1 | |
Query Syntax | TRIGger:SEQuence2:COUNt:CURRent? | |
| TRIGger:ACQuire:COUNt:CURRent? | |
Returned Parameters | <NR3> |
|
Related Commands | TRIG:SEQ2 | TRIG:ACQ |
TRIGger:SEQuence2:COUNt:VOLTage
TRIGger:ACQuire:COUNt:VOLTage
Agilent 66312A, 66332A Only
This command sets up a successive number of triggers for measuring voltage data. With this command, the trigger system needs to be initialized only once at the start of the acquisition period. After each completed measurement, the instrument waits for the next valid trigger condition to start another measurement. This continues until the count has completed.
Command Syntax | TRIGger:SEQuence2:COUNt:VOLTage<NRf+> | ||
| TRIGger:ACQuire:COUNt:VOLTage<NRf+> | ||
Parameters | 1 to 100 |
|
|
*RST Value | 1 |
|
|
Examples | TRIG:SEQ2:COUN:VOLT 5 | TRIG:ACQ:COUN:VOLT 1 | |
Query Syntax | TRIGger:SEQuence2:COUNt:VOLTage? | ||
| TRIGger:ACQuire:COUNt:VOLTage? | ||
Returned Parameters | <NR3> |
|
|
Related Commands | TRIG:SEQ2 | TRIG:ACQ |
|
75