Language Dictionary - 4
MEASure:VOLTage:LOW?
FETCh:VOLTage:LOW?
Agilent 66312A, 66332A Only
These queries return the Low level voltage of a voltage pulse waveform. The instrument first measures the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse waveform using 1024 bins between the maximum and minimum data points. The bin containing the most data points below the 50% point is the low bin. The average of all the data points in the low bin is returned as the Low level. If no low bin contains more than 1.25% of the total number of acquired points, then the minimum value is returned by these queries.
Query Syntax | MEASure[SCALar]:VOLTage:LOW? | |
| FETCh[:SCALar]:VOLTage:LOW? | |
Parameters | None |
|
Examples | MEAS:VOLT:LOW? | FETC:VOLT:LOW? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:VOLT:HIGH? | CALC:REF:LOW |
MEASure:VOLTage:MAXimum?
FETCh:VOLTage:MAXimum?
Agilent 66312A, 66332A Only
These queries return the maximum output voltage reading from the measurement sample.
Query Syntax | MEASure[:SCALar]:VOLTage:MAXimum? | |
| FETCh[:SCALar]:VOLTage:MAXimum? | |
Parameters | None |
|
Examples | MEAS:VOLT:MAX? | FETC:VOLT:MAX? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:VOLT:MIN? |
|
MEASure:VOLTage:MINimum?
FETCh:VOLTage:MINimum?
Agilent 66312A, 66332A Only
These queries return the minimum output voltage reading from the measurement sample.
Query Syntax | MEASure[:SCALar]:VOLTage:MINimum? | |
| FETCh[:SCALar]:VOLTage:MINimum? | |
Parameters | None |
|
Examples | MEAS:VOLT:MIN? | FETC:VOLT:MIN? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:VOLT:MAX? |
|
51