Agilent Technologies 6614C, 6634B SENSe WINDow, HANNing, RECTangular, Returned Parameters CRD

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 54
Image 54

4 - Language Dictionary

SENSe:WINDow

This command sets the window function that is used in output measurement calculations. The following functions can be selected:

HANNing

A signal conditioning window that reduces errors in dc and rms measurement

 

calculations in the presence of periodic signals such as line ripple. It also

 

reduces jitter when measuring successive pulse waveforms. The Hanning

 

window multiplies each point in the measurement sample by the function cos4.

 

Do not use the Hanning window when measuring single-shot pulse waveforms.

RECTangular

A window that returns measurement calculations without any signal

 

conditioning. This window may be used for pulse measurements where the

 

exact period of the pulse waveform is known and the measurement interval can

 

be set accordingly using the SENSe:SWEep:TINTerval command.

 

 

NOTE:

Neither window function alters the instantaneous voltage or current data returned in the

 

measurement array.

 

 

Command Syntax SENSe:WINDow[:TYPE] <type> Parameters HANNing RECTangular *RST Value HANNing

Examples SENS:WIND RECT Query Syntax SENSe:WINDow[:TYPE]?

Returned Parameters <CRD>

54