Language Dictionary - 4
TRIGger:SEQuence2:SOURce
TRIGger:ACQuire:SOURce
Agilent 66312A, 66332A Only
These commands select the trigger source for measurement triggers as follows:
BUS | GPIB device, *TRG, or <GET> (Group Execute Trigger) |
INTernal | trigger is generated internally when the measured waveform crosses the trigger level |
| with the selected slope. |
Command Syntax | TRIGger:SEQuence2:SOURce<source> |
| TRIGger:ACQuire:SOURce<source> |
Parameters | BUS INTernal |
*RST Value | INTernal |
Examples | TRIG:ACQ:SOUR BUS |
Query Syntax | TRIGger:SEQuence2:SOURce? |
| TRIGger:ACQuire:SOURce? |
Returned Parameters | <CRD> |
TRIGger:SEQuence1:DEFine
TRIGger:SEQuence2:DEFine
TRIGger:SEQuence2:DEFine applies to Agilent 66312A, 66332A Only
These commands define the names that are aliased to trigger sequences 1 and 2. The command accepts only ACQuire for sequence 2 and TRANsient for sequence 1 as predefined names. The query allows the user to query the instrument names aliased to sequences 1 and 2.
Command Syntax | TRIGger:SEQuence1:DEFine TRANsient | |
| TRIGger:SEQuence2:DEFine ACQuire | |
Parameters | TRANsient, ACQuire |
|
Examples | SEQ1:DEF ACQ | SEQ2:DEF TRAN |
Query Syntax | TRIGger:SEQuence1:DEFine? | |
| TRIGger:SEQuence2:DEFine? | |
Returned Parameters | <CRD> |
|
Related Commands | TRIG:SEQ2:ACQ | TRIG:SEQ1:TRAN |
*TRG
This common command generates a trigger when the trigger subsystem has BUS selected as its source. The command has the same affect as the Group Execute Trigger (<GET>) command.
In
Command Syntax | *TRG |
Parameters | None |
Related Commands | ABOR INIT TRIG[:IMM] <GET> |
79