4 - Language Dictionary
TRIGger:SEQuence2:HYSTeresis:CURRent
TRIGger:ACQuire:HYSTeresis:CURRent
Agilent 66312A, 66332A Only
This command defines a band around the trigger level through which the signal must pass before an internal measurement can occur. The band limit above and below the trigger level is one half of the hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an output waveform in the negative direction must start above the upper hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax | TRIGger:SEQuence2:HYSTeresis:CURRent<NRf+> | |
| TRIGger:ACQuire:HYSTeresis:CURRent<NRf+> | |
Parameters | 0 to MAX (see table |
|
Unit | A (amperes) |
|
*RST Value | 0 |
|
Examples | TRIG:SEQ2:HYST:CURR 0.5 TRIG:ACQ:HYST:CURR 0.5 | |
Query Syntax | TRIGger:SEQuence2:HYSTeresis:CURRent? | |
| TRIGger:ACQuire:HYSTeresis:CURRent? | |
Returned Parameters | <NR3> |
|
Related Commands | TRIG:SEQ2:HYST:VOLT | TRIG:SEQ2:LEV:CURR |
TRIGger:SEQuence2:HYSTeresis:VOLTage
TRIGger:ACQuire:HYSTeresis:VOLTage
Agilent 66312A, 66332A Only
This command defines a band around the trigger level through which the signal must pass before an internal measurement can occur. The band limit above and below the trigger level is one half of the hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an output waveform in the negative direction must start above the upper hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax | TRIGger:SEQuence2:HYSTeresis:VOLTage<NRf+> | |
| TRIGger:ACQuire:HYSTeresis:VOLTage<NRf+> | |
Parameters | 0 to MAX (see table |
|
Unit | V (volts) |
|
*RST Value | 0 |
|
Examples | TRIG:SEQ2:HYST:VOLT 2 | TRIG:ACQ:HYST:VOLT 2 |
Query Syntax | TRIGger:SEQuence2:HYSTeresis:VOLTage? | |
| TRIGger:ACQuire:HYSTeresis:VOLTage? | |
Returned Parameters | <NR3> |
|
Related Commands | TRIG:SEQ2:HYST:CURR | TRIG:SEQ2:LEV:VOLT |
76