Agilent Technologies 6631B, 6634B, 66332A, 6633B, 6632B, 6614C TRIGger SEQuence2 HYSTeresis CURRent

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 76
Image 76

4 - Language Dictionary

TRIGger:SEQuence2:HYSTeresis:CURRent

TRIGger:ACQuire:HYSTeresis:CURRent

Agilent 66312A, 66332A Only

This command defines a band around the trigger level through which the signal must pass before an internal measurement can occur. The band limit above and below the trigger level is one half of the hysteresis value added to or subtracted from the trigger level.

For a positive trigger to occur, the excursion of an output waveform in the positive direction must start below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an output waveform in the negative direction must start above the upper hysteresis band limit and pass through the lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:HYSTeresis:CURRent<NRf+>

 

TRIGger:ACQuire:HYSTeresis:CURRent<NRf+>

Parameters

0 to MAX (see table 4-3)

 

Unit

A (amperes)

 

*RST Value

0

 

Examples

TRIG:SEQ2:HYST:CURR 0.5 TRIG:ACQ:HYST:CURR 0.5

Query Syntax

TRIGger:SEQuence2:HYSTeresis:CURRent?

 

TRIGger:ACQuire:HYSTeresis:CURRent?

Returned Parameters

<NR3>

 

Related Commands

TRIG:SEQ2:HYST:VOLT

TRIG:SEQ2:LEV:CURR

TRIGger:SEQuence2:HYSTeresis:VOLTage

TRIGger:ACQuire:HYSTeresis:VOLTage

Agilent 66312A, 66332A Only

This command defines a band around the trigger level through which the signal must pass before an internal measurement can occur. The band limit above and below the trigger level is one half of the hysteresis value added to or subtracted from the trigger level.

For a positive trigger to occur, the excursion of an output waveform in the positive direction must start below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an output waveform in the negative direction must start above the upper hysteresis band limit and pass through the lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:HYSTeresis:VOLTage<NRf+>

 

TRIGger:ACQuire:HYSTeresis:VOLTage<NRf+>

Parameters

0 to MAX (see table 4-3)

 

Unit

V (volts)

 

*RST Value

0

 

Examples

TRIG:SEQ2:HYST:VOLT 2

TRIG:ACQ:HYST:VOLT 2

Query Syntax

TRIGger:SEQuence2:HYSTeresis:VOLTage?

 

TRIGger:ACQuire:HYSTeresis:VOLTage?

Returned Parameters

<NR3>

 

Related Commands

TRIG:SEQ2:HYST:CURR

TRIG:SEQ2:LEV:VOLT

76