4 - Language Dictionary
MEASure:VOLTage?
FETCh:VOLTage?
FETCh:VOLTage? applies to Agilent 66312A, 66332A Only
These queries return the dc output voltage.
Query Syntax | MEASure[:SCALar]:VOLTage[:DC]? | |
| MEASure[:SCALar]:VOLTage[:DC]? | |
Parameters | None |
|
Examples | MEAS:VOLT? | FETC:VOLT:DC? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:CURR? |
|
MEASure:VOLTage:ACDC?
FETCh:VOLTage:ACDC?
Agilent 66312A, 66332A Only
These queries return the ac+dc rms output voltage.
Query Syntax | MEASure[:SCALar]:VOLTage:ACDC? | |
| FETCh[:SCALar]:VOLTage:ACDC? | |
Parameters | None |
|
Examples | MEAS:VOLT:ACDC? | FETC:VOLT:ACDC? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:CURR:ACDC? |
|
MEASure:VOLTage:HIGH?
FETCh:VOLTage:HIGH?
Agilent 66312A, 66332A Only
These queries return the High level voltage of a voltage pulse waveform. The instrument first measures the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse waveform using 1024 bins between the maximum and minimum data points. The bin containing the most data points above the 50% point is the high bin. The average of all the data points in the high bin is returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points, then the maximum value is returned by these queries.
Query Syntax | MEASure[SCALar]:VOLTage:HIGH? | |
| FETCh[:SCALar]:VOLTage:HIGH? | |
Parameters | None |
|
Examples | MEAS:VOLT:HIGH? | FETC:VOLT:HIGH? |
Returned Parameters | <NR3> |
|
Related Commands | MEAS:VOLT:LOW? | CALC:REF:HIGH |