Agilent Technologies 6634B, 66332A, 6633B, 6632B, 6614C, 6613C, 6631B MEASure VOLTage? FETCh VOLTage?

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 50
Image 50

4 - Language Dictionary

MEASure:VOLTage?

FETCh:VOLTage?

FETCh:VOLTage? applies to Agilent 66312A, 66332A Only

These queries return the dc output voltage.

Query Syntax

MEASure[:SCALar]:VOLTage[:DC]?

 

MEASure[:SCALar]:VOLTage[:DC]?

Parameters

None

 

Examples

MEAS:VOLT?

FETC:VOLT:DC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR?

 

MEASure:VOLTage:ACDC?

FETCh:VOLTage:ACDC?

Agilent 66312A, 66332A Only

These queries return the ac+dc rms output voltage.

Query Syntax

MEASure[:SCALar]:VOLTage:ACDC?

 

FETCh[:SCALar]:VOLTage:ACDC?

Parameters

None

 

Examples

MEAS:VOLT:ACDC?

FETC:VOLT:ACDC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:ACDC?

 

MEASure:VOLTage:HIGH?

FETCh:VOLTage:HIGH?

Agilent 66312A, 66332A Only

These queries return the High level voltage of a voltage pulse waveform. The instrument first measures the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse waveform using 1024 bins between the maximum and minimum data points. The bin containing the most data points above the 50% point is the high bin. The average of all the data points in the high bin is returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points, then the maximum value is returned by these queries.

Query Syntax

MEASure[SCALar]:VOLTage:HIGH?

 

FETCh[:SCALar]:VOLTage:HIGH?

Parameters

None

 

Examples

MEAS:VOLT:HIGH?

FETC:VOLT:HIGH?

Returned Parameters

<NR3>

 

Related Commands

MEAS:VOLT:LOW?

CALC:REF:HIGH