Agilent Technologies 6612C, 6634B, 66332A, 6633B, 6632B, 6614C, 6613C MEASure CURRent? FETCh CURRent?

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 48
Image 48

4 - Language Dictionary

MEASure:CURRent?

FETCh:CURRent?

FETCh:CURRent? applies to Agilent 66312A, 66332A Only

These queries return the dc output current.

Query Syntax

MEASure:[SCALar]:CURRent[:DC]?

 

FETCh:[SCALar]:CURRent[:DC]?

Parameters

None

Examples

MEAS:CURR? MEAS:CURR:DC?

Returned Parameters

<NR3>

Related Commands

MEAS:VOLT?

MEASure:CURRent:ACDC?

FETCh:CURRent:ACDC?

Agilent 66312A, 66332A Only

These queries return the ac+dc rms output current.

Query Syntax

MEASure:[SCALar]:CURRent:ACDC?

 

FETCh:[SCALar]:CURRent:ACDC?

Parameters

None

 

Examples

MEAS:CURR:ACDC?

FETC:CURR:ACDC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:VOLT:ACDC?

 

MEASure:CURRent:HIGH?

FETCh:CURRent:HIGH?

Agilent 66312A, 66332A Only

These queries return the High level current of a current pulse waveform. The instrument first measures the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse waveform using 1024 bins between the maximum and minimum data points. The bin containing the most data points above the 50% point is the high bin. The average of all the data points in the high bin is returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points, then the maximum value is returned by these queries.

Query Syntax

MEASure[SCALar]:CURRent:HIGH?

 

FETCh[:SCALar]:CURRent:HIGH?

Parameters

None

Examples

MEAS:CURR:HIGH?FETC:CURR:HIGH?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR:LOW? CALC:REF:HIGH

48

Page 48
Image 48
Agilent Technologies 6612C, 6634B, 66332A, 6633B MEASure CURRent? FETCh CURRent?, MEASure CURRent ACDC? FETCh CURRent ACDC?