Agilent Technologies 6614C Calibration Commands, CALibrate CURRent NEGative, Cal Curr Meas

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 44
Image 44
Calibration Commands

4 - Language Dictionary

Calibration Commands

Calibration commands let you:

υEnable and disable the calibration mode

υChange the calibration password

υCalibrate the current and voltage programming and measurement, and store new calibration constants in nonvolatile memory.

NOTE: If calibration mode has not been enabled with CALibrate:STATe, programming the calibration commands will generate an error.

CALibrate:CURRent

This command initiates the calibration of the positive dc output current as well as the high-range current measurement circuit.

Command Syntax

CALibrate:CURRent[:SOURce][:DC][:POSitive]

Parameters

None

Examples

CAL:CURRCAL:CURR:SOUR:DC:POS

Related Commands

CAL:CURR:NEG

CALibrate:CURRent:NEGative

This command initiates the calibration of the negative dc output current.

Command Syntax

CALibrate:CURRent[:SOURce][:DC]:NEGative

Parameters

None

Examples

CAL:CURR:NEGCAL:CURR:SOUR:DC:NEG

Related Commands

CAL:CURR

CALibrate:CURRent:MEASure:LOWRange

This command initiates the calibration of the low-range current measurement circuit.

Command Syntax

CALibrate:CURRent:MEASure[:DC]:LOWRange

Parameters

None

Examples

CAL:CURR:MEAS

Related Commands

CAL:CURR

CALibrate:CURRent:MEASure:AC

Agilent 66312A, 66332A Only

This command initiates the calibration of the high bandwidth (ac) measurement circuit.

Command Syntax

CALibrate:CURRent:MEASure:AC

Parameters

None

Examples

CAL:CURR:MEAS:AC

44

Page 44
Image 44