Language Dictionary - 4
SENSe:FUNCtion
Agilent 66312A, 66332A Only
This command configures the measurement sensor to measure either voltage or current when an acquire trigger is used. The query returns the function setting, either VOLT or CURR.
Command Syntax SENSe:FUNCtion <function
Parameters "VOLTage" "CURRent"
Examples SENS:FUNC "VOLT"
Query Syntax SENSe:FUNCtion?
Returned Parameters <SRD>
SENSe:SWEep:OFFSet:POINts
Agilent 66312A, 66332A Only
This command defines the offset in a data sweep when an acquire trigger is used. Negative values represent data samples taken prior to the trigger.
Command Syntax | SENSe:SWEep:OFFSet:POINts <NRf+> |
Parameters | |
*RST Value | 0 |
Examples | SENS:SWE:OFFS:POIN |
Query Syntax | SENSe:SWEep:OFFSet:POINts? |
Returned Parameters | <NR3> |
Related Commands | SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR |
SENSe:SWEep:POINts
This command defines the number of points in a data sweep.
Command Syntax | SENSe:SWEep:POINts<NRf+> |
Parameters | 0 through 4096 |
*RST Value | 2048 |
Examples | SENS:SWE:POIN 1024 |
Query Syntax | SENSe:SWEep:POINts? |
Returned Parameters | <NR3> |
Related Commands | SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR |
SENSe:SWEep:TINTerval
This command defines the time period between samples
Command Syntax | SENSe:SWEep:TINTerval<NRf+> |
Parameters | 15.6 microseconds through 31200 seconds |
*RST Value | 15.6 microseconds |
Examples | SENS:SWE:TINT |
Query Syntax | SENSe:SWEep:TINTerval? |
Returned Parameters | <NR3> |
Related Commands | SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR |
53