OUTPut:RELay

Agilent 66332A, 6632B, 6633B, 6634B, 6611C, 6612C, 6613C, 6614C Only

This command is only valid for units with Relay Option 760, otherwise an error will occur. Programming ON closes the output relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the dc source is supplying power to a load, that power will appear at the relay contacts during switching.

Command Syntax

OUTPut:RELay[:STATe]<bool>

Parameters

0

1 OFF ON

*RST Value

0

 

Examples

OUTP:REL 1OUTP:REL OFF

Query Syntax

OUTPput:RELay?

Returned Parameters

0

1

Related Commands

OUTP *RCL *SAV

OUTPut:RELay:POLarity

Agilent 66332A, 6632B, 6633B, 6634B, 6611C, 6612C, 6613C, 6614C Only

This command is only valid for units with Relay Option 760, otherwise an error will occur. Programming NORMal causes the output relay polarity to be the same as the dc source output. Programming REVerse causes the relay output polarity to be opposite to that of the dc source output. If OUTPut = ON when either command is sent, the output voltage is set to 0 during the time that the relays are changing polarity.

Command Syntax

OUTPut:RELay:POLarity<CRD>

Parameters

NORMal REVerse

*RST Value

NORM

Examples

OUTP:REL:POL NORM

Query Syntax

OUTPput:RELay:POLarity?

Returned Parameters

NORM REV

Related Commands

OUTP *RCL *SAV

OUTPut:RI:MODE

This command selects the mode of operation of the Remote Inhibit protection. The RI mode is stored in non-volatile memory. The following modes can be selected:

LATChing

causes a TTL low signal on the INH input to disable the output. The only way

 

to clear the latch is by sending an OUTPut:PROTection:CLEAR command

 

while the INH input is false.

LIVE

allows the INH input to disable the output in a non-latching manner. In other

 

words, the output follows the state of the INH input. When INH is low true, the

 

output is disabled. When INH is high the output is not affected.

OFF

the INH input is disabled.

Command Syntax

Parameters

Examples

Query Syntax

Returned Parameters

Related Commands

OUTPut:RI:MODE <mode> LATChing LIVE OFF OUTP:RI:MODE LIVE OUTPut:RI:MODE? <CRD>

OUTP:PROT:CLE