Agilent Technologies 6612C, 6634B TRIGger SEQuence2 SLOPe CURRent, TRIGger ACQuire SLOPe CURRent

Models: 6632B 6633B 66332A 6634B

1 103
Download 103 pages 34 Kb
Page 78
Image 78
POSitive NEGative EITHer

4 - Language Dictionary

TRIGger:SEQuence2:SLOPe:CURRent

TRIGger:ACQuire:SLOPe:CURRent

Agilent 66312A, 66332A Only

This command sets the slope of an internally triggered current measurement.

POSitive

triggering occurs on the rising edge.

NEGative

triggering occurs on the falling edge.

EITHer

triggering occurs on either edge.

Command Syntax

TRIGger:SEQuence2:SLOPe:CURRent<slope>

 

TRIGger:ACQuire:SLOPe:CURRent<slope>

Parameters

EITHerPOSitiveNEGative

 

*RST Value

EITHer

 

Examples

TRIG:SEQ2:SLOP:CURR POS

TRIG:ACQ:SLOP:CURR EITH

Query Syntax

TRIGger:SEQuence2:SLOPe:CURRent?

 

TRIGger:ACQuire:SLOPe:CURRent?

Returned Parameters

<CRD>

 

Related Commands

TRIG:SEQ2:SLOP:VOLT

 

TRIGger:SEQuence2:SLOPe:VOLTage

TRIGger:ACQuire:SLOPe:VOLTage

Agilent 66312A, 66332A Only

This command sets the slope of an internally triggered voltage measurement. triggering occurs on the rising edge.

triggering occurs on the falling edge. triggering occurs on either edge.

Command Syntax

TRIGger:SEQuence2:SLOPe:VOLTage<slope>

 

TRIGger:ACQuire:SLOPe:VOLTage<slope>

Parameters

EITHerPOSitiveNEGative

*RST Value

EITHer

Examples

TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH

Query Syntax

TRIGger:SEQuence2:SLOPe:VOLTage?

 

TRIGger:ACQuire:SLOPe:VOLTage?

Returned Parameters

<CRD>

Related Commands

TRIG:SEQ2:SLOP:CURR

78

Page 78
Image 78
Agilent Technologies 6612C TRIGger SEQuence2 SLOPe CURRent, TRIGger ACQuire SLOPe CURRent, TRIGger SEQuence2 SLOPe VOLTage