Status commands program the dc source status registers. The dc source has three groups of status registers; Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as described later in this section. The Operation and Questionable status groups each consist of the Condition, Enable, and Event registers and the NTR and PTR filters. Chapter 3 under "Programming the Status Registers" explains how to read specific register bits and use the information they return.

Common commands also perform status functions. The following common commands are discussed in

this section: *CLS *ESE *SR? *OPC *PSC *SRE *STB *WAI.

STATus:PRESet

This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsytem NTR and Enable registers.

Command Syntax

STATus:PRESet

Parameters

None

Examples

STAT:PRES STATUS:PRESET

Bit Position

Bit Name

Bit Weight

Table 4-4. Bit Configuration of Operation Status Registers

15–12

11

10

9

8

7-6

5

not

CC-

CC+

not

CV

not

WTG

used

 

 

used

 

used

 

 

2048

1024

 

256

 

32

4-1

not

used

0

CAL

1

CAL = The dc source is computing new calibration constants.

WTG = The dc source is waiting for a trigger.

CV = The dc source is operating in constant voltage mode.

CC+ = The dc source is operating in constant current mode.

CC = The dc source is operating in negative constant current mode.

STATus:OPERation?

This query returns the value of the Operation Event register. The Event register is a read-only register which holds (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.

Query Syntax

STATus:OPERtion[:EVENt]?

Parameters

None

 

Returned Parameters

<NR1>(Register Value)

 

Examples

STAT:OPER?

STATUS:OPERATIONAL:EVENT?

Related Commands

*CLS STAT:OPER:NTR STAT:OPER:PTR

STATus:OPERation:CONDition?

This query returns the value of the Operation Condition register. That is a read-only register which holds the real-time (unlatched) operational status of the dc source .

Query Syntax

STATus:OPERation:CONDition?

Parameters

None

 

Examples

STAT:OPER:COND?

STATUS:OPERATION:CONDITION?

Returned Parameters

<NR1> (register value)

 