Commands by Subsystem 3

:EXTernal:BWLimit

 

(see page 530)

Command Syntax

:EXTernal:BWLimit <bwlimit>

 

<bwlimit> ::= {0 OFF}

 

The :EXTernal:BWLimit command is provided for product compatibility.

 

The only legal value is 0 or OFF. Use the :TRIGger:HFReject command to

 

limit bandwidth on the external trigger input.

Query Syntax

:EXTernal:BWLimit?

 

The :EXTernal:BWLimit? query returns the current setting of the low- pass

 

filter (always 0).

Return Format

<bwlimit><NL>

 

<bwlimit> ::= 0

See Also

"Introduction to :EXTernal Trigger Commands" on page 171

 

"Introduction to :TRIGger Commands" on page 313

 

":TRIGger:HFReject" on page 317

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

173