248 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:MEASure:VBASe

(see page 530)
Command Syntax :MEASure:VBASe [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VBASe command installs a screen measurement and starts a
waveform base value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax :MEASure:VBASe? [<source>]
The :MEASure:VBASe? query returns the vertical value at the base of the
waveform. The base value of a pulse is normally not the same as the
minimum value.
Return Format <base_voltage><NL>
<base_voltage> ::= value at the base of the selected waveform in
NR3 format
See Also "Introduction to :MEASure Commands" on page 218
":MEASure:SOURce" on page 240
":MEASure:VTOP" on page 254
":MEASure:VAMPlitude" on page 246
":MEASure:VMIN" on page 250
NOTE This command is not available if the source is FFT (Fast Fourier Transform).