3Commands by Subsystem

:MEASure:VBASe

(see page 530)

Command Syntax :MEASure:VBASe [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:VBASe command installs a screen measurement and starts a waveform base value measurement. If the optional source parameter is specified, the current source is modified.

N O T E

This command is not available if the source is FFT (Fast Fourier Transform).

 

Query Syntax :MEASure:VBASe? [<source>]

The :MEASure:VBASe? query returns the vertical value at the base of the waveform. The base value of a pulse is normally not the same as the minimum value.

Return Format <base_voltage><NL>

<base_voltage> ::= value at the base of the selected waveform in NR3 format

See Also "Introduction to :MEASure Commands" on page 218

":MEASure:SOURce" on page 240

":MEASure:VTOP" on page 254

":MEASure:VAMPlitude" on page 246

":MEASure:VMIN" on page 250