Commands by Subsystem 3

:CHANnel<n>:PROBe:STYPe

 

(see page 530)

Command Syntax

This command is valid only for the 113xA Series probes.

N O T E

 

:CHANnel<n>:PROBe:STYPe <signal type> <signal type> ::= {DIFFerential SINGle}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :CHANnel<n>:PROBe:STYPe command sets the channel probe signal type (STYPe) to differential or single- ended when using the 113xA Series probes and determines how offset is applied.

When single- ended is selected, the :CHANnel<n>:OFFset command changes the offset value of the probe amplifier. When differential is selected, the :CHANnel<n>:OFFset command changes the offset value of the channel amplifier.

Query Syntax :CHANnel<n>:PROBe:STYPe?

The :CHANnel<n>:PROBe:STYPe? query returns the current probe signal type setting for the selected channel.

Return Format <signal type><NL>

<signal type> ::= {DIFF SING}

See Also "Introduction to :CHANnel<n> Commands" on page 143

":CHANnel<n>:OFFSet" on page 151

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference