3Commands by Subsystem

:CHANnel<n>:PROBe

(see page 530)

Command Syntax :CHANnel<n>:PROBe <attenuation>

<attenuation> ::= probe attenuation ratio in NR3 format

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The obsolete attenuation values X1, X10, X20, X100 are also supported.

The :CHANnel<n>:PROBe command specifies the probe attenuation factor for the selected channel. The probe attenuation factor may be 0.1 to 1000. This command does not change the actual input sensitivity of the oscilloscope. It changes the reference constants for scaling the display factors, for making automatic measurements, and for setting trigger levels.

If an AutoProbe probe is connected to the oscilloscope, the attenuation value cannot be changed from the sensed value. Attempting to set the oscilloscope to an attenuation value other than the sensed value produces an error.

Query Syntax :CHANnel<n>:PROBe?

The :CHANnel<n>:PROBe? query returns the current probe attenuation factor for the selected channel.

Return Format <attenuation><NL>

<attenuation> ::= probe attenuation ratio in NR3 format

See Also "Introduction to :CHANnel<n> Commands" on page 143

":CHANnel<n>:RANGe" on page 157

":CHANnel<n>:SCALe" on page 158

":CHANnel<n>:OFFSet" on page 151

Example Code ' CHANNEL_PROBE - Sets the probe attenuation factor for the selected

'channel. The probe attenuation factor may be set from 0.1 to 1000. myScope.WriteString ":CHAN1:PROBE 10" ' Set Probe to 10:1.

Example program from the start: "VISA COM Example in Visual Basic" on page 578

152

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 152
Image 152
Agilent Technologies 5000 Series manual CHANnelnPROBe