Agilent Technologies 5000 Series manual Aver Hres

Models: 5000 Series

1 598
Download 598 pages 46.18 Kb
Page 54
Image 54

2Commands Quick Reference

Table 27 :WAVeform Commands Summary (continued)

Command

Query

Options and Query Returns

 

 

 

:WAVeform:POINts

:WAVeform:POINts?

<# points> ::= {100 250 500

<# points> (see

(see page 413)

1000 <points_mode>} if waveform

page 413)

 

points mode is NORMal

 

 

<# points> ::= {100 250 500

 

 

1000 2000 ... 8000000 in 1-2-5

 

 

sequence <points_mode>} if

 

 

waveform points mode is MAXimum

 

 

or RAW

 

 

<points_mode> ::= {NORMal

 

 

MAXimum RAW}

 

 

 

:WAVeform:POINts:MODE

:WAVeform:POINts:MODE

<points_mode> ::= {NORMal

<points_mode> (see

? (see page 415)

MAXimum RAW}

page 415)

 

 

 

 

 

n/a

:WAVeform:PREamble?

<preamble_block> ::= <format

 

(see page 417)

NR1>, <type NR1>,<points

 

 

NR1>,<count NR1>, <xincrement

 

 

NR3>, <xorigin NR3>, <xreference

 

 

NR1>,<yincrement NR3>, <yorigin

 

 

NR3>, <yreference NR1>

 

 

<format> ::= an integer in NR1

 

 

format:

 

 

• 0 for BYTE format

 

 

• 1 for WORD format

 

 

• 2 for ASCii format

 

 

<type> ::= an integer in NR1

 

 

format:

 

 

• 0 for NORMal type

 

 

• 1 for PEAK detect type

 

 

• 2 for AVERage type

 

 

• 3 for HRESolution type

 

 

<count> ::= Average count, or 1

 

 

if PEAK detect type or NORMal; an

 

 

integer in NR1 format

 

 

 

:WAVeform:SOURce

:WAVeform:SOURce?

<source> ::= {CHANnel<n>

<source> (see

(see page 420)

FUNCtion MATH}

page 420)

 

<n> ::= 1-2 or 1-4 in NR1 format

 

 

 

:WAVeform:SOURce:SUBS

:WAVeform:SOURce:SUBS

<subsource> ::= {{NONE RX}

ource <subsource>

ource? (see page 424)

TX}

(see page 424)

 

 

 

 

 

n/a

:WAVeform:TYPE? (see

<return_mode> ::= {NORM PEAK

 

page 425)

AVER HRES}

 

 

 

54

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 54
Image 54
Agilent Technologies 5000 Series manual Aver Hres