Commands by Subsystem 3

:FUNCtion:CENTer

 

(see page 530)

Command Syntax

:FUNCtion:CENTer <frequency>

 

<frequency> ::= the current center frequency in NR3 format. The range

 

of legal values is from 0 Hz to 25 GHz.

 

The :FUNCtion:CENTer command sets the center frequency when FFT

 

(Fast Fourier Transform) is selected.

Query Syntax

:FUNCtion:CENTer?

 

The :FUNCtion:CENTer? query returns the current center frequency in

 

Hertz.

Return Format

<frequency><NL>

 

<frequency> ::= the current center frequency in NR3 format. The range

 

of legal values is from 0 Hz to 25 GHz.

 

After a *RST (Reset) or :AUToscale command, the values returned by the

N O T E

:FUNCtion:CENTer? and :FUNCtion:SPAN? queries depend on the current :TIMebase:RANGe

 

 

value. Once you change either the :FUNCtion:CENTer or :FUNCtion:SPAN value, they no

 

longer track the :TIMebase:RANGe value.

 

 

See Also "Introduction to :FUNCtion Commands" on page 182

":FUNCtion:SPAN" on page 191

":TIMebase:RANGe" on page 306

":TIMebase:SCALe" on page 308

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

183

Page 183
Image 183
Agilent Technologies 5000 Series manual FUNCtionCENTer, Longer track the TIMebaseRANGe value