184 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:FUNCtion:DISPlay

(see page 530)
Command Syntax :FUNCtion:DISPlay <display>
<display> ::= {{1 | ON} | {0 | OFF}}
The :FUNCtion:DISPlay command turns the display of the function on or
off. When ON is selected, the function performs as specified using the
other FUNCtion commands. When OFF is selected, function is neither
calculated nor displayed.
Query Syntax :FUNCtion:DISPlay?
The :FUNCtion:DISPlay? query returns whether the function display is on
or off.
Return Format <display><NL>
<display> ::= {1 | 0}
See Also "Introduction to :FUNCtion Commands" on page 182
":VIEW" on page 122
":BLANk" on page 94
":STATus" on page 119