Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 179

:EXTernal:RANGe

(see page 530)
Command Syntax :EXTernal:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :EXTernal:RANGe command is provided for product compatibility. The
range can only be set to 5.0 V when using 1:1 probe attenuation. If the
probe attenuation is changed, the range value is multiplied by the probe
attenuation factor.
Query Syntax :EXTernal:RANGe?
The :EXTernal:RANGe? query returns the current full- scale range setting
for the external trigger.
Return Format <range_argument><NL>
<range_argument> ::= external trigger range value in NR3 format = (5.0 V
) * (probe attenuation factor)
See Also "Introduction to :EXTernal Trigger Commands" on page 171
":EXTernal:PROBe" on page 175
"Introduction to :TRIGger Commands" on page 313