Commands by Subsystem 3

:EXTernal:PROBe:STYPe

 

(see page 530)

Command Syntax

This command is valid only for the 113xA Series probes.

N O T E

 

:EXTernal:PROBe:STYPe <signal type>

<signal type> ::= {DIFFerential SINGle}

The :EXTernal:PROBe:STYPe command sets the external trigger probe signal type (STYPe) to differential or single- ended when using the 113xA Series probes and determines how offset is applied.

Query Syntax :EXTernal:PROBe:STYPe?

The :EXTernal:PROBe:STYPe? query returns the current probe signal type setting for the external trigger.

Return Format <signal type><NL>

<signal type> ::= {DIFF SING}

See Also "Introduction to :EXTernal Trigger Commands" on page 171

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

177

Page 177
Image 177
Agilent Technologies 5000 Series manual EXTernalPROBeSTYPe, Command Syntax