Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 249

:MEASure:VMAX

(see page 530)
Command Syntax :MEASure:VMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VMAX command installs a screen measurement and starts a
maximum vertical value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax :MEASure:VMAX? [<source>]
The :MEASure:VMAX? query measures and outputs the maximum vertical
value present on the selected waveform.
Return Format <value><NL>
<value> ::= maximum vertical value of the selected waveform in
NR3 format
See Also "Introduction to :MEASure Commands" on page 218
":MEASure:SOURce" on page 240
":MEASure:VMIN" on page 250
":MEASure:VPP" on page 251
":MEASure:VTOP" on page 254