5Obsolete and Discontinued Commands

:MEASure:LOWer

 

 

(see page 530)

Command Syntax

:MEASure:LOWer <voltage>

 

 

The :MEASure:LOWer command sets the lower measurement threshold

 

 

value. This value and the UPPer value represent absolute values when the

 

 

thresholds are ABSolute and percentage when the thresholds are PERCent

 

 

as defined by the :MEASure:DEFine THResholds command.

 

 

The :MEASure:LOWer command is obsolete and is provided for backward compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 222)

 

 

 

 

instead.

 

 

Query Syntax

:MEASure:LOWer?

 

 

The :MEASure:LOWer? query returns the current lower threshold level.

Return Format

<voltage><NL>

 

 

<voltage> ::= the user-defined lower threshold in volts in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

 

":MEASure:THResholds" on page 481

 

 

":MEASure:UPPer" on page 488

478

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 478
Image 478
Agilent Technologies 5000 Series manual MEASureLOWer, Instead