Obsolete and Discontinued Commands 5

:MEASure:VSTOp

 

 

(see page 530)

Command Syntax

:MEASure:VSTOp <vstop_argument>

 

 

<vstop_argument> ::= value for Y2 cursor

 

 

The :MEASure:VSTOp command moves the vertical marker 2 (Y2 cursor) to

 

 

the specified value corresponding to the selected source. The source can

 

 

be selected by the MARKer:X2Y2source command.

 

 

The short form of this command, VSTO, does not follow the defined Long Form to Short

N O T E

 

 

Form Truncation Rules (see page 532). The normal short form, VST, would be the same for

 

 

 

 

both VSTArt and VSTOp, so sending VST for the VSTOp command produces an error.

 

 

 

 

 

The :MEASure:VSTOp command is an obsolete command provided for compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MARKer:Y2Position command (see page 212) instead.

 

 

 

 

Query Syntax

:MEASure:VSTOp?

 

 

The :MEASure:VSTOp? query returns the current value of the Y2 cursor.

Return Format

<value><NL>

 

 

<value> ::= value of the Y2 cursor in NR3 format

See Also

"Introduction to :MARKer Commands" on page 204

 

 

"Introduction to :MEASure Commands" on page 218

 

 

":MARKer:Y1Position" on page 211

 

 

":MARKer:Y2Position" on page 212

 

 

":MARKer:YDELta" on page 213

 

 

":MARKer:X2Y2source" on page 209

 

 

":MEASure:SOURce" on page 240

 

 

":MEASure:TDELta" on page 480

 

 

":MEASure:TSTArt" on page 484

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

491

Page 491
Image 491