:TRIGger:LIN:SIGNal:DEFinition

Command Syntax

N O T E

(see page 530)

:TRIGger:LIN:SIGNal:DEFinition <value>

<value> ::= {LIN RX TX}

The :TRIGger:LIN:SIGNal:DEFinition command sets the LIN signal type. These signals can be set to:

Dominant low signals:

LIN — the actual LIN single- end bus signal line.

RX — the Receive signal from the LIN bus transceiver.

TX — the Transmit signal to the LIN bus transceiver.

With InfiniiVision 5000 Series oscilloscope software version 5.00 or greater, this command is available, but the only legal value is LIN.

Query Syntax

:TRIGger:LIN:SIGNal:DEFinition?

 

The :TRIGger:LIN:SIGNal:DEFinition? query returns the current LIN signal

 

type.

Return Format

<value><NL>

 

<value> ::= LIN

See Also

"Introduction to :TRIGger Commands" on page 313

 

":TRIGger:MODE" on page 319

 

":TRIGger:LIN:SIGNal:BAUDrate" on page 367

 

":TRIGger:LIN:SOURce" on page 368

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference